EBT2/3 Film for Auto QA Tool  - CyberKnife

EBT2/3 Film for Auto QA Tool (100 pieces)

Product SKU: 022354-002

The automated quality assurance (AQA) test is a targeting reproducibility test. The AQA test should be performed daily to check robot mastering of the treatment manipulator and stability of the imaging system. The AQA test checks the reproducibility of targeting two beams (vertical and horizontal). The test delivers a 1-path, 2-node treatment plan to film. From film analysis results, the robot targeting and imaging system constancy can be determined and analyzed.

Each Kit Includes: 
Box of film, EBT2/3 (100 pieces)